Application of Fisher's electrophotometer to thiamine dosage

Authors

  • O. F. Ribeiro Universidade de São Paulo, Escola Politécnica, São Paulo, SP
  • Lucia Aché Universidade de São Paulo, Faculdade de Medicina Veterinária e Zootecnia, Departamento de Química Orgânica e Biológica, São Paulo, SP

DOI:

https://doi.org/10.11606/issn.2318-5066.v3i1/2p31-33

Keywords:

The article has no key-words.

Abstract

Utilising the Fisher electrophotometer for the dosing of tiamin it was thereby verified to be necessary the concomitant fact of a standard tube which must be utilizes as if it were the blank. The dosings realized by the electrophotometer with a quantity of tiamin, varying from 30 to 200 by ml do not show errors higher than one per cent.